Acquisition material
There are different setup to acquire the iEEG with the epileptic patients, using 2 different amplfiers: a clinical one and a reserch one
Characteristics of the amplifiers
Clinical amplifiers
Nicolet (< 2010)
This amplifier has been used from xxxx to xxxx in order to record macroelectrodes
- maximum number of channels :
- samplig rate : 400 Hz
- input range :
- bandpass
Micromed: SD LTM 64 BS (2010 - July 2020)
This amplifier is used since xxxx in order to record macroelectrodes.
- maximum number of channels : 128 - 2 boxes of 64 inputs each
- samplig rate : 1024 Hz
- input range : +/- 3.2 mV
- bandpass : 0.18 - 220 Hz
Micromed: SD LTM 64 PLUS (September 2020 - )
This amplifier is used since September 2020 (pat_02968_1239) in order to record macroelectrodes.
- maximum number of channels : 128 - 2 boxes of 64 inputs each
- samplig rate : 1024 Hz
- input range : +/- 3.2 mV
- bandpass : 0.18 - 220 Hz
Research amplifiers
Neuralynx : Lynx (2010 - July 2012)
This amplifier has been used from 2010 to July 2010 in order to record only microelectrodes, with Cheetah software
Neuralynx : ATLAS (November 2012 - )
This amplifier is used since October 2012 in order to record microelectrodes and macroelectrodes.
- maximum number of channels : 160
- samplig rate : flexible but set to 4kHz for the macroelectrdoes and 32 kHz for the microelectrodes
- input range : flexible but set to +/- 4mV
- bandpass : flexible but set to 0.1 - 1000 Hz for the macroelectrodes and 0.1 - 8000 Hz for the microelectrodes
- acquisition software :
- Cheetah : October 2012 - Mai 2022 (pat_02006_0436 - pat_03152_1614)
- march 2022: direct RJ45 cable between Neuralynx and Micromed PCs (pat_03138_1601)
- Pegasus : June 2022 - (pat_03167_1630 - )
- Cheetah : October 2012 - Mai 2022 (pat_02006_0436 - pat_03152_1614)
Electrodes
Macroelectrodes
The macroelectrodes (AdTech®, Wisconsin) consist of 4-12 platinum contact electrodes with a diameter of 1.12 mm and length of 2.41 mm, with nickel-chromium wiring. The distance between the center of 2 contacts is 5 mm.
Microelectrodes
The macro-microelectrodes (AdTech®, Wisconsin) are of the Behnke-Fried type, consisting of 8 platinum contact electrodes with a diameter of 1.28 mm and length of 1.57 mm, embedded on the surface of a polyurethane tube with a hollow lumen. Eight 40 µm platinum-iridium microwires were inserted into the macroelectrode, protruding 3-6 mm into the cerebral tissue beyond the tip of the macroelectrode. The protruding microwires were cut to the desired length during the surgical procedure. The distance between contact 1 and 2 is 3 mm, between 2 and 3, 3 and 4, etc is 7 mm.
Connectors
Adtech CABRIO connectors
Those connectors are used for macroelectrodes
Neuralynx Headstages CHETs
Those connectors are used for microelectrodes, and can also be used for macroelectrodes
Acquisition setups (- macroelectrodes / -- microelectrodes)
setup #1: Clinical setup
amplifier:
- Micromed
electrodes:
- macroelectrodes
connectors:
- AdTech
period used
setup #2: Macro and micro electrodes separated
amplifier:
- Micromed
- Neuralynx
electrodes:
- macroelectrodes to Micromed
- microelectrodes to Neuralynx
connectors:
- CHETs for all electrodes
period used:
setup #3: Macro and micro electrodes together, but 2 amplifiers in parallel
amplifier:
- Micromed
- Neuralynx
electrodes:
- macroelectrodes to both systems
- microelectrodes at least to Neuralynx
connectors:
- CHETs for all electrodes
period used:
- start in October 2012
- end in Apr 2018
- + 1 patient in november 2018
setup #4: Macro and micro electrodes together, only 1 amplifier, using Neuralynx CHETs
amplifier:
- Neuralynx (sent to Micromed PC)
electrodes:
- macroelectrodes to both acquisition PC
- microelectrodes to both acquisition PC
connectors:
- CHETs for all electrodes
- TopHat adapters beginning of Apr 2018 (end of recording) and July 2018 (whole recording)
period used:
- tests March 2018 and June 2018 (whole recording with Nlx//MM in parallel) and July 2018 (whole recording)
- october 2018
setup #5: Macro and micro electrodes together, only 1 amplifier, using AdTech connectors
amplifier:
- Neuralynx (sent to Micromed PC)
electrodes:
- macroelectrodes to both acquisition PC
- microelectrodes to both acquisition PC
connectors:
- AdTech for macroeletrodes + Neuralynx Headboxes
- CHETs for microeletrodes
period used:
- tests Jan 2019
- January 2019- (pat_02680_1158-)
Sleep electrodes
Since february 2017 (pat_02459_0912) scalp electrodes were added in order to record sleep activity. Some of the electrodes (emg and E1, E2) are removed during the week ends
- O1, Oz, O2
- E1, E2
- C3, M2 or C4, M1 for right and left implantation respectively
For the analysis they should be rereferenced as follow
Updates
September 2020
The epilepsy unit moved from Paul Castaigne to Marguerite Bottard.
The Micromed system has been renewed at the same time, but the Neuralynx system remains the same.
The first patient recorded in Margurite Bottard is pat_02968_1239, with the new Micromed system only
March 2022
Direct RJ45 cable installed between Neuralynx and Micromed PCs (first patient: pat_03138_1601)
June 2022
Upgrade of the Neuralynx software from Cheetah to Pegasus and update of Micromed's software.
First patient recorded with the new woftwares: pat_03167_1630
December 2023
The epilepsy moved back to the renew building now called : Institut de Neurologie
First patient recorded : pat_03364_1840