Depending on the acquisition settings, the analysis requier different preprocessing steps.
Nlx2MM
This configuration is to be avoided.
TTL are sent to Neuralynx that send them to Micromed:
- TTL numbers are all presents and with correct number
- Timings of TTL on Neuralynx are good
- Timings of TTL on Micromed have up to 200 ms delay
Solution: take timing from neurlaynx and find the time dely between Neuralynx and Micromed to transpose the correct TTL timings to Micromed
Neuralynx and Micromed in parallel
...